Device Troubleshooting
Practice describing device behavior, separating sensing from capture problems, and escalating findings precisely.
Learning objectives
The device troubleshooting framework
Pacing artifact without expected QRS
Describe this as an output followed by no obvious expected ventricular capture on the available tracing. Confirm the tracing and escalate for formal device evaluation rather than changing programming independently.
Intrinsic QRS appears, but device timing behaves as if it was not detected
This raises a sensing question. Correlate surface ECG with the device electrogram/markers during formal interrogation and consider whether the intrinsic event met the system's sensing criteria.
Device records very rapid events that do not match the surface rhythm
Do not assume true tachyarrhythmia. Stored electrograms and marker channels are needed to evaluate possibilities such as oversensing or other signal-related behavior.
Better communication
Instead of “the pacemaker isn't working,” report the observable finding: “There is a ventricular pacing artifact without an obvious subsequent QRS on this tracing.”
What is the best first troubleshooting step?