Activation Maps & LAT
Activation mapping is fundamentally a timing problem: every mapped point is interpreted relative to a reference and an annotation.
Learning objectives
Window of interest
The mapping system displays timing within a selected window around the reference. The window should match the electrophysiologic question. If the reference moves or the rhythm changes, points acquired under different conditions may no longer be comparable.
When automatic annotation struggles
Fractionated signals, double potentials, far-field components, noise, poor contact and changing electrogram morphology can cause an algorithm to select a component that does not represent the intended local activation.
The “earliest” point has a noisy electrogram
Inspect the raw signal, neighboring bipoles, catheter stability, contact information when available, reference timing and nearby points. Do not accept the color alone as proof of earliest activation.
An activation map suddenly becomes physiologically confusing. What should be checked early?